Device product
SIGNA Architect
Z-2130-2025
Product summary
- Event
- Event 97030
- Status
- Ongoing
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-2130-2025
Official wording
Reason: GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Code information: GTIN: 00840682147095 00840682122702 00195278023643 00840682123440
Distribution pattern: Worldwide distribution - US Nationwide and the country of China.
Derived failure modes
-
Unknown
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.