Device product
3mensio Workstation (Vascular Fenestrated) software
Z-2297-2025
Product summary
- Event
- Event 97307
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 501
- Official record key
device-enforcement:Z-2297-2025
Official wording
Reason: When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage
Code information: Software Version/UDI/Software Release Date: 10.6/(01)08056304455505(11)240325(8012)10.6/05-Apr-24, 10.7/(01)08056304455925(11)250423(8012)10.7/29-Apr-25, Service Packs/UDI: 10.6 SP1/(01)08056304455505(11)240411(8012)10.6 SP1/23-Apr-24, 10.6 SP2/(01)08056304455505(11)240610(8012)10.6 SP2/17-Jun-24, 10.6 SP3/(01)08056304455505(11)240926(8012)10.6 SP3/03-Oct-24, 10.6 SP4/(01)08056304455505(11)250109(8012)10.6 SP4/20-Jan-25, 10.7 SP1/(01)08056304455925(11)250523(8012)10.7 SP1/28-May-25
Distribution pattern: US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.
Derived failure modes
-
Unknown
When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage