Device product
LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model: NCTK31L All convenience kits contain Specialist Blend Cast Padding, Item: QTX30-322NS
Z-1916-2025
Product summary
- Event
- Event 96837
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 850 Kits
- Official record key
device-enforcement:Z-1916-2025
Official wording
Reason: Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.
Code information: Model/UDI-DI/Lot(Expiration): NCEX20I/00191072222398/9146304(2025-12-03), 9207685(2026-01-08); NCEX20J/00191072234582/9222096(2026-01-30), 9222096(2026-01-30); NCTK31K/00191072222480/9086225(2025-11-01), 9086225(2025-11-01); NCTK31L/00191072234872/9219596(2026-01-15), 9219596(2026-01-15)
Distribution pattern: US Nationwide distribution in the state of CO.
Derived failure modes
-
Unknown
Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.