Skip to content
Recall Observatory FDA recall evidence

Device product

PhenoMATRIX

Z-1812-2025

May 06, 2025

Class II

Product summary

Firm
Copan WASP
Event
Event 96765
Status
Ongoing
Classification
Class II
Quantity
14
Official record key
device-enforcement:Z-1812-2025

Official wording

Reason: AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.

Code information: All software versions

Distribution pattern: US: Nationwide distribution in the states of UT, AR, CA, PA, MD, NC, IL, MA, NJ, KS, TX, NY, NV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.