Device product
Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
Z-2198-2025
Product summary
- Firm
- Capnia Inc
- Event
- Event 97184
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 2160
- Official record key
device-enforcement:Z-2198-2025
Official wording
Reason: Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
Code information: Monitor UDI-DI: 00858481006017. Lot (C20206): 231220
Distribution pattern: US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.
Derived failure modes
-
Unknown
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.