Device product

VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.

Z-2073-2025

June 12, 2025

Class II

Product summary

Firm
Ortho-Clinical Diagnostics, Inc.
Event
Event 97034
Status
Ongoing
Classification
Class II
Quantity
2,017 units
Official record key
device-enforcement:Z-2073-2025

Official wording

Reason: Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.

Code information: Model Number: 680 1997. UDI-DI: 10758750001835. Lot Numbers: 1527-15-2749 (exp. 17-Nov-2025), 1527-15-2771 (exp. 17-Dec-2025), 1527-15-2818 (exp. 11-Apr-2026), 1527-15-2871 (exp. 15-May-2026).

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.