Device product
iGo¿2 DC Car Adapter to be used with iGo¿2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.
Z-1727-2025
Product summary
- Event
- Event 96700
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 89,532 units
- Official record key
device-enforcement:Z-1727-2025
Official wording
Reason: Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.
Code information: DC Cord Code Info: Catalog No. DV6X-619; UDI 885304020585; Cord Revision: Rev E; Lot No. beginning with 2023-01-01 through 2024-12-31. iGo2 Portable Oxygen Concentrator Systems Code Info: Catalog No. 125D, 125D-XB, 125D-BT, 125D-BT-XB, 125D-ARYA-XB; UDI 885304022237, 885304032250, 885304033257, 885304033240, 885304033516; Lot No. D23125001DS to D23A26121DS, D23706001DA to D23B16056DA, F23B06001DS to F25108143DS, F23313001DH to F25109259DH, F24212001DA to F25217150DA.
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Albania, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, ISRAEL, Italy, JAPAN, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Maldives, Nepal, Netherlands, Norway, Oman, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovak Republic, Slovenia, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad & Tobago, Turkey, Uganda, Ukraine, United Arab Emirates.
Derived failure modes
-
Unknown
Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.