Device product
TufTex Over-the-Wire Embolectomy Catheter. All Catalog numbers: 1651-34, 1651-38, 1651-44, 1651-48, 1651-58, 1651-64, 1651-68, 1651-78, 1651-84, 1651-88, E1651-34, E1651-38, E1651-44, E1651-48, E1651-64, E1651-68, E1651-78, E1651-84, E1651-88, 1651-44J, 1651-48J, 1651-84J, 1651-88J.
Z-1750-2025
Product summary
- Event
- Event 96716
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 237,611 units
- Official record key
device-enforcement:Z-1750-2025
Official wording
Reason: During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).
Code information: UDI-DI: 00840663100651, 00840663100668, 00840663100675, 00840663100682, 00840663100699, 00840663100705, 00840663100712, 00840663100729, 00840663107551, 00840663100736, 00840663100743, 00840663100750, 00840663100767, 00840663100774, 00840663100781, 00840663100798, 00840663100804, 00840663100811, 00840663100828, 00840663100835, 00840663107568, 00840663100842, 00840663100859, 00840663100866. Catalog numbers: 1651-34, 1651-38, 1651-44, 1651-48, 1651-68, 1651-78, 1651-84, 1651-88, E1651-34, E1651-38, E1651-44, E1651-48, E1651-64, E1651-68, E1651-78, E1651-84, E1651-88. Lots not expired with the following prefix: OTW, QOT, XOT, COT
Distribution pattern: Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CAMBODIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECHIA, DENMARK, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONGOLIA, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORTH MACEDONIA, NORWAY, OMAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKIYE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.
Derived failure modes
-
Sterility assurance
packages to have incomplete seals (sterile