Maquet Getinge-BEQ-TOP 38302 INFANT CPS QPID Material: 701055866

Z-1543-2019

April 18, 2019

Class II

Product summary

Firm
Maquet Cardiovascular, LLC
Event
Event 82703
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-1543-2019

Official wording

Reason: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Code information: Batch Numbers: 3000045221 3000049201 3000054750 3000062046 3000065497 3000067793 3000069723 3000072247 3000075339 3000080720 3000084636

Distribution pattern: Nationwide

Derived failure modes

  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    Sterility Assurance Level (SAL) of the stopcock ports cannot be assured