Device product
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
Z-1665-2025
Product summary
- Event
- Event 96559
- Status
- Ongoing
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-1665-2025
Official wording
Reason: The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
Code information: All Serial Numbers. 08N53-001, UDI-DI: 00884999047389; 08N53-002, UDI-DI: 00884999048034; 08N53-032, UDI-DI: 00884999047587
Distribution pattern: Worldwide - US Nationwide distribution.
Derived failure modes
-
Unknown
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.