Skip to content
Recall Observatory FDA recall evidence

Device product

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No

Z-1654-2025

March 18, 2025

Class II

Product summary

Firm
Abbott Laboratories
Event
Event 96567
Status
Ongoing
Classification
Class II
Quantity
19
Official record key
device-enforcement:Z-1654-2025

Official wording

Reason: When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Code information: UDI-DI 00380740017170 SNs: 35405BG 36034BG 36259BG 36326BG 36327BG 36328BG 36372BG 36545BG 36556BG 36583BG 36588BG 36589BG 54539BG 55361BG 55454BG 55520BG 55814BG 55828BG 72229BG

Distribution pattern: Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.