Skip to content
Recall Observatory FDA recall evidence

Device product

Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Number: TFL-SLS, TFL-SLS-JA (Japan only); Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: No

Z-1920-2025

May 06, 2025

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 96723
Status
Ongoing
Classification
Class II
Quantity
201 units (100 US, 101 OUS)
Official record key
device-enforcement:Z-1920-2025

Official wording

Reason: Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

Code information: All lot/serial numbers for the following products: TFL-SLS UDI: 00821925044135; TFL-SLS-JA (Japan Only)

Distribution pattern: Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.