Skip to content
Recall Observatory FDA recall evidence

Device product

Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber Component: No. a conduit for respiratory gas between the patient and a ventilator.

Z-1698-2025

April 10, 2025

Class I

Product summary

Firm
Vyaire Medical
Event
Event 96664
Status
Ongoing
Classification
Class I
Quantity
3053
Official record key
device-enforcement:Z-1698-2025

Official wording

Reason: Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.

Code information: Lot Code: Lot/Serial Number(s): 0004247924 0004247925 0004252940 0004252941 0004256292 0004256293 0004256294 0004256295 0004272325 0004272868 0004278978 0004279156 0004280036 0004285304 0004288036 0004289514 UDI-DI: AirLife Label Each: 10889483595909 Case: 30889483595903 Vyaire Label: Each: 10190752145160 Case: 50190752145168

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Germany, Poland, Belgium, Spain, Romania, UK, Italy, Netherlands, India, Slovenia, Qatar, and Saudi Arabia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.