Skip to content
Recall Observatory FDA recall evidence

Device product

CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)

Z-1269-2025

January 09, 2025

Class II

Product summary

Firm
Kung Shin Plastics Co. Ltd.
Event
Event 96224
Status
Ongoing
Classification
Class II
Quantity
30000 ea/ 300 cases
Official record key
device-enforcement:Z-1269-2025

Official wording

Reason: The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.

Code information: UDI (01)00810050853682(10)C2N24002(11)240828(17)290827 Lot C2N24002

Distribution pattern: US Nationwide distribution in the state of IL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.