Skip to content
Recall Observatory FDA recall evidence

Device product

A.L.P.S. mvX-MEDIAL TIBIA PLATE 10H RT-Medial Tibia Plate, 10-Hole, Right Intended Fixation of fractures of the distal tibia. Model/Catalog Number: 770715102

Z-1222-2025

January 21, 2025

Class II

Product summary

Firm
Tyber Medical
Event
Event 96212
Status
Ongoing
Classification
Class II
Quantity
209 units
Official record key
device-enforcement:Z-1222-2025

Official wording

Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Code information: UDI-DI 00196449016006; Lot Numbers 265815, 24055DF03, 24055DF04

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices