Device product
NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01RF NIM 4.0 REFURBISHED, REF: NIM4CM01RF; PATIENT INTERFACE NIM4CPB1 NIM 4.0, REF: NIM4CPB1; PATIENT INTFC NIM4CPB1RF NIM 4.0 REFURB, REF: NIM4CPB1RF; SOFTWARE NIM4SWU143 UPGRADE V1.4.3, REF: NIM4SWU143; SOFTWARE NIM4SWU154 UPGRADE V1.5.4, REF: NIM4SWU154; SOFTWARE NIM4SWU164 UPGRADE V1.6.4, REF: NIM4SWU164
Z-1623-2025
Product summary
- Event
- Event 96511
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 9,788
- Official record key
device-enforcement:Z-1623-2025
Official wording
Reason: Nerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead to procedure delay, extubation, and medical intervention.
Code information: All NIM Vital Nerve Monitoring Systems manufactured or installed with software version v1.6.4 or earlier. REF/UDI-DI: NIM4CM01/763000002978, 00763000395896, 00763000401597, 00763000528577; NIM4CM01RF/00763000002992; NIM4CPB1/763000002985, 00763000401603, 00763000395902, 00763000528584; NIM4CPB1RF/00763000003005; NIM4SWU143/00763000709341, 00763000869823; NIM4SWU154/00763000945398; NIM4SWU164/00763000974312
Distribution pattern: Worldwide - US Nationwide distribution including in the states of MD, MN, PA, OK, OH, FL, CO, GA, NC, KS, IL, WY, AK, NE, VA, MI, CA, NY, WI, UT, AZ, AR, TX, WA, AL, NJ, SD, MS, KY, TN, MO, OR, MT, MA, CT, NM, LA, WV, NH, DC, DE, IN, NV, IA, SC, ID, ME, HI, RI, VT, ND, GU, PR and the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Namibia, Nepal, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
Derived failure modes
-
Device software or design
software versions has potential for increased stimulus artifact, which may require troubleshooting, and may