Device product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer
Z-1664-2025
Product summary
- Event
- Event 96554
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 370 units
- Official record key
device-enforcement:Z-1664-2025
Official wording
Reason: When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
Code information: UDI/DI 15099590732103, all units run with system software versions SW 1.20.0 and below
Distribution pattern: Worldwide distribution.
Derived failure modes
-
Unknown
When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.