Device product
Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235; in-vitro diagnostic test
Z-1164-2025
Product summary
- Event
- Event 96166
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 27 units
- Official record key
device-enforcement:Z-1164-2025
Official wording
Reason: Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.
Code information: UDI/DI 70845357043077, Lot Numbers: 8235-32, 8235-33
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.
Derived failure modes
-
Unknown
Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.