Device product
IntelePACS (Image Fusion Module) - InteleViewer
Z-1208-2025
Product summary
- Event
- Event 96270
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 88
- Official record key
device-enforcement:Z-1208-2025
Official wording
Reason: A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Code information: UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later
Distribution pattern: Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
Derived failure modes
-
Device software or design
bug in specific software