Device product

IntelePACS (Image Fusion Module) - InteleViewer

Z-1208-2025

February 05, 2025

Class II

Product summary

Firm
INTELERAD MEDICAL SYSTEMS INCORPORATED
Event
Event 96270
Status
Ongoing
Classification
Class II
Quantity
88
Official record key
device-enforcement:Z-1208-2025

Official wording

Reason: A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

Code information: UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later

Distribution pattern: Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    bug in specific software