Skip to content
Recall Observatory FDA recall evidence

Device product

A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 10H RT- Anatomic Lateral Fibula Plate, 10-Hole, Right. Intended forFixation of fractures of the distal tibia Model/Catalog Number: 770708102 Product Description:

Z-1216-2025

January 21, 2025

Class II

Product summary

Firm
Tyber Medical
Event
Event 96212
Status
Ongoing
Classification
Class II
Quantity
219 units
Official record key
device-enforcement:Z-1216-2025

Official wording

Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Code information: UDI-DI 00196449015597; Lot Numbers 265810, 260892, 262559

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices