Skip to content
Recall Observatory FDA recall evidence

Device product

HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553570

Z-1037-2025

December 20, 2024

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 96049
Status
Ongoing
Classification
Class II
Quantity
118 units (OUS)
Official record key
device-enforcement:Z-1037-2025

Official wording

Reason: Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.

Code information: GTIN: 08714729976684 Lot Numbers: 32878539, 33336847, 33338446, 33775901, 33795273, 33900261, 33900262, 34076575, 34083270, 34172108, 34172109

Distribution pattern: Nationwide including Puerto Rico Foreign: To be provided

Derived failure modes

  • Unknown