Device product
Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz
Z-1901-2018
Product summary
- Firm
- LivaNova USA
- Event
- Event 80072
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2408
- Official record key
device-enforcement:Z-1901-2018
Official wording
Reason: Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.
Code information: Serial #: a) 16S10743 - 6S11708 b) 16S10772 - 16S16840 C) 16S10958 - 16S16847
Distribution pattern: MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,
Derived failure modes
-
Microbial contamination
microbial contamination