Device product
Allon 2001. Part Number: 200-00263
Z-1966-2025
Product summary
- Event
- Event 96854
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 229 units
- Official record key
device-enforcement:Z-1966-2025
Official wording
Reason: An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.
Code information: Part Number: 200-00263. UDI-DI: 007290107581103, Serial Numbers: 99111143000002, 99111143000014, 99111627300034, 9993901600001, 9993901600002, 9993901600004, 9993901600006, 9993901600007, 9993901600009, 9995356300005, 9995356300006, 9995356300007, 9995356300008, 9996673800002, 99111143000001, 99111143000003, 99111143000004, 99111143000007, 99111143000008, 99111143000009, 99111143000011, 99111143000012, 99111143000013, 99111143000015, 99111143000018, 99111143000019, 99111288700016, 99111288700021, 99111288700035, 99111288700036, 99111627300011, 99111627300017, 99111627300022, 99111627300029, 99111627300040, 99111627300041, 99111627300045, 99111627300046, 99111627300054, 99111627300057, 99111627300064, 99111627300067, 99113168300004, 99113168300005, 99113168300008, 99113168300010, 99113168300012, 99128623400008, 99129614700002, 99129614700005, 99131046400001, 99131046400003, 99131046400004, 99131046400005, 99131046400006, 99131046400009, 99131046400011, 99131046500007, 99132732400009, 99132732400018, 99132732400019, 99134232100011, 99132121000010, 99113168300003, 9983349800008, 9991451200002, 9991451200004, 9992175000001, 9995356300010, 99111627300001, 99111627300008, 99113168300013, 99132121000014, 99132732400005, 99134232100001, 99134232100002, 99134232100003, 99134232100009, 99134232100022, 99134232100035, 99111627300042, 99111627300056, 99111627300058, 99111627300059, 99113168300014, 99111627300020, 99111627300049, 99113168300006, 99128623400003, 99132121000013, 99134232100018, 9991451200001, 99111143000006, 99111627300005, 99111627300035, 99113168300001, 99113168300002, 99113168300007, 99111627300053, 99111288700014, 99111288700017, 99111288700018, 99111288700019, 99111288700020, 99111627300016, 99111627300032, 99111143000017, 99111143000021, 99111143000022, 99111288700001, 99111288700002, 99129614700009, 9962076700011, 99132121000012, 99132121000015, 99132121000005, 99132121000007, 99111627300009, 99134232100017, 99131046400008, 9995356100003, 99113168300011, 99111627300025, 99132121000009, 99111627300061, 99111627300038, 9974083900001, 99111627300030, 9991451200006, 99111627300018, 9976534100006, 9945767700003, 9983349700008, 9991451200003, 9991451200009, 99111288700005, 99111627300031, 99128623400004, 99128623400005, 99128623400007, 99129614700001, 99129614700003, 99129614700004, 99129614700007, 99129614700008, 99131046400002, 99131046500001, 99131046500002, 99132121000001, 99128623400006, 99111143000016, 9969414000006, 99134232100013, 99134232100021, 99111627300036, 99132121000017, 99132121000018, 99132121000019, 99132121000020, 99132732400002, 99132732400006, 99132732400013, 99132732400014, 99134232100005, 99134232100010, 99132121000016, 99111627300013, 99111627300033, 99111627300043, 99111627300044, 99111627300051, 99111627300063, 99129614700006, 99111627300007, 99111627300024, 99111627300037, 99111627300047, 99111627300014, 99111627300052, 99113168300009, 99128623400001, 99128623400002, 99128623400009, 99128623400010, 99134232100008, 99111143000020, 99131046400010, 99132121000011, 99132732400007, 99132732400011, 99132732400012, 99132732400016, 99134232100034, 99132732400008, 99132732400015, 99132732400017, 99132732400020, 99134232100014, 9959060500014, 99111627300026, 99134232100004, ", 99134232100006", 9991451200005, 9995356200008, 99111143000010, 99111288700004, 99131046500003, 99131046500005, 9973536800007, 99131046500004, 99131046500008, 99131046500009, 99132121000004, 99132121000006, 99132121000008, 99134232100026, 99132732400001, 99132732400003, 99132732400004, 99134232100007, 99134232100019, 9940577300005, 99111143000005, 99111627300048, 99111627300062, 99132121000002, 9979841500003, 99132121000003, 99131046500010
Distribution pattern: Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.
Derived failure modes
-
Unknown
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.