Device product
Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062
Z-1070-2025
Product summary
- Event
- Event 96097
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 2,939
- Official record key
device-enforcement:Z-1070-2025
Official wording
Reason: Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.
Code information: UDI-DI: REF/UDI-DI/Batch: 154150041/00863328000189/332905 - 349049, 154150042/328641, REF: 154090042/322101, 156135061/00863328000103/332824 - 336121, 206135061/00850003494043/331281 - 336152, 206135062/328643 - 345567, 156135062/271330 - 283066, 206080062/321083 - 349035, 156080062/302251 - 349034
Distribution pattern: Distribution US nationwide and countries of: NL, DE, NL, AU, IT, CH, GB, and AT
Derived failure modes
-
Unknown
Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.