Skip to content
Recall Observatory FDA recall evidence

Device product

Signature Laparoscopic Instruments, Debakey Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5862100.

Z-1895-2025

January 09, 2024

Class II

Product summary

Firm
Silex Medical, LLC
Event
Event 96755
Status
Ongoing
Classification
Class II
Quantity
8 units
Official record key
device-enforcement:Z-1895-2025

Official wording

Reason: Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

Code information: Model/REF Number: 5862100. UDI-DI: B33158621000. Lot Number: M28183. Serial Numbers: 1932, 1933, 1934, 1935, 1936, 1937, 1938, 1939.

Distribution pattern: US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.