Device product
Access Testosterone assay, Catalog Number 33560
Z-1901-2025
Product summary
- Event
- Event 96824
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 70 units
- Official record key
device-enforcement:Z-1901-2025
Official wording
Reason: Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
Code information: UDI-DI: 15099590231347; Serial Number: 538081
Distribution pattern: US State: FL. UAE, Vietnam
Derived failure modes
-
Unknown
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.