Skip to content
Recall Observatory FDA recall evidence

Device product

Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.

Z-0905-2025

December 18, 2024

Class I

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 95898
Status
Ongoing
Classification
Class I
Quantity
30,140 units
Official record key
device-enforcement:Z-0905-2025

Official wording

Reason: Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.

Code information: Model Number: MAJ-891. UDI-DI: 04953170063114. All lot numbers.

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0