Skip to content
Recall Observatory FDA recall evidence

Device product

regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.

Z-1047-2025

December 26, 2024

Class I

Product summary

Firm
ROi CPS LLC
Event
Event 96094
Status
Ongoing
Classification
Class I
Quantity
414 kits
Official record key
device-enforcement:Z-1047-2025

Official wording

Reason: There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..

Code information: Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.

Distribution pattern: Distribution was made to MO. There was no government/military/foreign distribution.

Derived failure modes

  • Potency or specification failure

    reason.potency_specification_failure · v1.0.0

    out of specification