Skip to content
Recall Observatory FDA recall evidence

Device product

MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71

Z-0665-2025

November 28, 2023

Class II

Product summary

Firm
Micro-X Ltd.
Event
Event 95617
Status
Ongoing
Classification
Class II
Quantity
18
Official record key
device-enforcement:Z-0665-2025

Official wording

Reason: Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.

Code information: MXU-RV35 UDI-DI: 9357123000037 and MXU-RV71 UDI-DI: 935712300005 using Control Board PCBA, 12425-03 (or earlier). Serial Numbers: MSN00334, MSN00348, MSN00349, MSN00351, MSN00352, MSN00361, MSN00364, MSN00365,MSN00382, MSN00383, MSN00384, MSN00385, MSN00386, MSN00387, MSN00388, MSN00390, MSN00391, MSN00395

Distribution pattern: US Nationwide Distribution: NJ, CA, PR

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.