Device product
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStation 11A and 11A Service Pack 1, 2 and 3
Z-1189-2025
Product summary
- Event
- Event 96156
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 6
- Official record key
device-enforcement:Z-1189-2025
Official wording
Reason: Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
Code information: Lot Code: UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916 and 0735000201063120220616 GTIN: 07350002010389, 07350002010433, 07350002010440 and 07350002010631 Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStation 11A and 11A Service Pack 1, 2 and 3 Expiration Date: 2027-06-22
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK.
Derived failure modes
-
Unknown
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.