Device product

StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.

Z-0681-2025

November 20, 2024

Class I

Product summary

Firm
Nova Biomedical Corporation
Event
Event 95783
Status
Ongoing
Classification
Class I
Quantity
2,365 units
Official record key
device-enforcement:Z-0681-2025

Official wording

Reason: A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

Code information: Model No 63685. UDI-DI (01)10385480636858. Software Versions: v0.0.13.10 to v0.0.13.44

Distribution pattern: Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software bug