Device product
SURG STRP 1X6-200 Model/Catalog Number: 801453. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Z-0252-2025
Product summary
- Event
- Event 95432
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 10719
- Official record key
device-enforcement:Z-0252-2025
Official wording
Reason: Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
Code information: UDI-DI: 10381780515098, 20886704036613 Note: the same product code was distributed for a few years under JnJ under a transition agreement with Integra. Therefore, 2 UDI numbers are listed here for JnJ and for Integra. All unexpired products.
Distribution pattern: Worldwide Distribution. US Nationwide, Canada Germany United Kingdom of Great Britain and Northern Ireland (the) Malta Italy Barbados Trinidad and Tobago Australia New Zealand Netherlands (the) Belgium France Ireland Switzerland Brunei Darussalam Bahamas (the) South Africa China Ecuador Korea (the Republic of) Czechia Kuwait Luxembourg French Guiana Mayotte Burundi Reunion Martinique Poland Costa Rica Japan Senegal Turkey Serbia Mexico Cambodia Spain Ghana Iceland Jersey Iraq Bahrain Lebanon Dominican Republic (the) Singapore Portugal Greece Austria C¿te d'Ivoire Palestine, State of Georgia Nigeria Slovenia Sweden Colombia United Arab Emirates (the) Saudi Arabia Qatar Hong Kong Oman Albania Panama Holy See (the) Kenya Pakistan Guernsey Morocco Chile Brazil Argentina Puerto Rico Rwanda Jordan Philippines (the) Hungary Romania Bolivia (Plurinational State of) Malaysia El Salvador Peru Lithuania Republic of North Macedonia Taiwan (Province of China) Israel Cyprus Bulgaria
Derived failure modes
-
Potency or specification failure
out-of-specification