Device product
smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP
Z-0527-2025
Product summary
- Event
- Event 95605
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 142327 units
- Official record key
device-enforcement:Z-0527-2025
Official wording
Reason: Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
Code information: REF 14-504-17, UDI/DI 35021239002579, Lot Numbers: 0003340, 0003620, 0003616, 0003580, 0003542, 0003509, 0003579, 0003568, 0003599, 0003615, 0003589, 0003562, 0003601, 0003618, 0003583, 0003613, 0003270, 0003629, 0003293, 0003584, 0003487, 0003617, 0003280, 0003348, 0003470, 0003523, 0003607, 0003576, 0003422, 0003614, 0003633, 0003578, 0003630, 0003207, 0003438, 0003559, 0003570, 0003478, 0003533, 0003588, 0003433, 0003556, 0003561, 0003446, 0003284, 0003545, 0003461, 0003590, 0003398, 0003530, 0003539, 0003550, 0003213, 0003598, 0003623, 0003504, 0003548, 0003552, 0003631, 0003594, 0003625, 0003600, 0003378, 0003526, 0003574, 0003411, 0003471, 0003587, 0003596, 0003297, 0003298, 0003412, 0003424, 0003425, 0003517, 0003151, 0003163, 0003235, 0003347, 0003350, 0003382, 0003390, 0003416, 0003423, 0003603, 0003638, 0003641, 0003175, 0003214, 0003238, 0003253, 0003257, 0003343, 0003351, 0003388, 0003397, 0003428, 0003486, 0003488, 0003496, 0003498, 0003528, 0003541, 0003608, 0003610, 0003612, 0003639, 0003640, 0003121, 0003122, 0003123, 0003125, 0003126, 0003127, 0003128, 0003129, 0003131, 0003132, 0003133, 0003134, 0003135, 0003136, 0003137, 0003138, 0003141, 0003142, 0003143, 0003145, 0003146, 0003147, 0003149, 0003150, 0003153, 0003154, 0003155, 0003156, 0003157, 0003158, 0003159, 0003160, 0003164, 0003166, 0003167, 0003168, 0003170, 0003171, 0003172, 0003173, 0003174, 0003176, 0003177, 0003178, 0003179, 0003180, 0003181, 0003182, 0003183, 0003188, 0003190, 0003191, 0003192, 0003193, 0003194, 0003195, 0003197, 0003198, 0003199, 0003200, 0003201, 0003202, 0003203, 0003208, 0003209, 0003211, 0003212, 0003215, 0003217, 0003219, 0003220, 0003221, 0003223, 0003224, 0003225, 0003227, 0003228, 0003229, 0003230, 0003231, 0003233, 0003234, 0003237, 0003239, 0003240, 0003241, 0003242, 0003243, 0003244, 0003245, 0003246, 0003247, 0003248, 0003249, 0003250, 0003251, 0003252, 0003258, 0003259, 0003260, 0003261, 0003262, 0003263, 0003264, 0003265, 0003266, 0003268, 0003269, 0003271, 0003272, 0003273, 0003274, 0003275, 0003278, 0003279, 0003281, 0003282, 0003283, 0003285, 0003286, 0003287, 0003288, 0003289, 0003290, 0003291, 0003292, 0003294, 0003295, 0003296, 0003299, 0003302, 0003303, 0003304, 0003305, 0003306, 0003309, 0003310, 0003311, 0003312, 0003313, 0003316, 0003317, 0003318, 0003319, 0003320, 0003321, 0003323, 0003324, 0003325, 0003326, 0003327, 0003328, 0003330, 0003331, 0003332, 0003333, 0003334, 0003335, 0003336, 0003337, 0003338, 0003341, 0003342, 0003344, 0003345, 0003346, 0003349, 0003352, 0003353, 0003354, 0003355, 0003356, 0003365, 0003366, 0003379, 0003383, 0003384, 0003385, 0003386, 0003387, 0003389, 0003392, 0003396, 0003400, 0003401, 0003402, 0003404, 0003405, 0003406, 0003407, 0003408, 0003409, 0003410, 0003413, 0003417, 0003418, 0003419, 0003420, 0003421, 0003426, 0003427, 0003429, 0003430, 0003431, 0003432, 0003434, 0003435, 0003436, 0003437, 0003439, 0003440, 0003441, 0003442, 0003443, 0003444, 0003445, 0003447, 0003448, 0003450, 0003451, 0003452, 0003453, 0003454, 0003455, 0003458, 0003460, 0003462, 0003463, 0003464, 0003465, 0003466, 0003467, 0003468, 0003472, 0003473, 0003474, 0003475, 0003479, 0003481, 0003482, 0003485, 0003490, 0003491, 0003493, 0003494, 0003495, 0003497, 0003499, 0003501, 0003502, 0003503, 0003505, 0003506, 0003507, 0003508, 0003512, 0003513, 0003514, 0003515, 0003516, 0003518, 0003519, 0003520, 0003521, 0003524, 0003525, 0003527, 0003531, 0003532, 0003534, 0003535, 0003536, 0003537, 0003538, 0003543, 0003544, 0003547, 0003551, 0003554, 0003560, 0003565, 0003566, 0003582, 0003586, 0003591, 0003592, 0003593, 0003595, 0003602, 0003604, 0003606, 0003611, 0003624, 0003632, 0003636, 0003637, 0003642, 0003575, 0003540, 0003572, 0003558, 0003555, 0003567, 0003585, 0003546, 0003609, 0003564, 0003563, 0003571, 0003577, 0003569; REF 14-504-17JP, UDI/DI 35021239002579, Lot Numbers: 0003215, 0003415, 0003483
Distribution pattern: Worldwide distribution.
Derived failure modes
-
Unknown
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.