BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352

Z-0906-2025

April 05, 2023

Class II

Product summary

Firm
CareFusion 303, Inc.
Event
Event 96002
Status
Ongoing
Classification
Class II
Quantity
622
Official record key
device-enforcement:Z-0906-2025

Official wording

Reason: Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.

Code information: Software v1.7.3. REF/UDI-DI: 323/10885403512667, 352/10885403512674

Distribution pattern: US Nationwide distribution in the states of NY, NC, LA.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software issue