Skip to content
Recall Observatory FDA recall evidence

Device product

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Z-0664-2025

May 31, 2023

Class II

Product summary

Firm
Micro-X Ltd.
Event
Event 95603
Status
Ongoing
Classification
Class II
Quantity
5
Official record key
device-enforcement:Z-0664-2025

Official wording

Reason: Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Code information: UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Distribution pattern: Worldwide and US Nationwide Distribution: NJ, PR, CA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.