Device product

DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens

Z-0677-2025

November 07, 2024

Class II

Product summary

Firm
Diasorin Inc.
Event
Event 95829
Status
Ongoing
Classification
Class II
Quantity
21905 units
Official record key
device-enforcement:Z-0677-2025

Official wording

Reason: Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.

Code information: UDI/DI 80567713190605F, Lot Numbers: 225084, 232094, 234114, 259144, 221214, 252244, 230094, 224124, 205144, 228174, 228224, 223274, 219104, 217134, 233154, 236174, 224244, 232294, 229224, 251234, 221294, 210204, 223244, 222314

Distribution pattern: Worldwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.