Skip to content
Recall Observatory FDA recall evidence

Device product

O&M Halyard SOFT TOUCH II Surgical Mask, Blue-Disposable surgical mask Product Code: 47500

Z-0729-2025

November 08, 2024

Class II

Product summary

Firm
O&M HALYARD, INC.
Event
Event 95804
Status
Ongoing
Classification
Class II
Quantity
637 cases
Official record key
device-enforcement:Z-0729-2025

Official wording

Reason: Not meeting the bioburden testing requirements of EN 14683 required for masks sold in Europe. Face masks sold in the United States are not required to meet this standard. However, if the standard is printed on the packaging, compliance is required.

Code information: UDI-DI: 20680651475002 Lot Number: AM4119961 Exp. Date: 2029-04-28

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Not meeting the bioburden testing requirements of EN 14683 required for masks sold in Europe. Face masks sold in the United States are not required to meet this standard. However, if the standard is printed on the packaging, compliance is required.