Recall Observatory FDA recall evidence

Device product

NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10

Z-1039-2025

January 06, 2025

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 96090
Status
Ongoing
Classification
Class II
Quantity
155 units
Official record key
device-enforcement:Z-1039-2025

Official wording

Reason: It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.

Code information: UDI-DI: 00889024635746; Lot Numbers:66782843 66782840 66873137

Distribution pattern: Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.