Device product
Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each processing step. Siemens Material Number(SMN): 11099326 All lots starting with N15 or 15 are expected to have the defect.
Z-0496-2025
Product summary
- Event
- Event 95626
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 9729 units
- Official record key
device-enforcement:Z-0496-2025
Official wording
Reason: A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and patient results for seven (7) analytes: Ammonia, Acetaminophen, Alanine Aminotransferase (ALT), Hemoglobin A1c, Creatinine Kinase (CK), Cystatin C, and Salicylate.
Code information: Lot # UDI N1527024 (01)00630414596143(10)N1527024(17)20291002; N1526424 (01)00630414596143(10)N1526424(17)20291001; N1525924 (01)00630414596143(10)N1525924(17)20290924 ; N1525324 (01)00630414596143(10)N1525324(17)20290918 ; N1524224 (01)00630414596143(10)N1524224(17)20290911; N1523724 (01)00630414596143(10)N1523724(17)20290904; N1523024 (01)00630414596143(10)N1523024(17)20290827; N1521824 (01)00630414596143(10)N1521824(17)20290814; N1521424 (01)00630414596143(10)N1521424(17)20290808; N1519324 (01)00630414596143(10)N1519324(17)20290719; N1518924 (01)00630414596143(10)N1518924(17)20290716; N1518224 (01)00630414596143(10)N1518224(17)20290710; N1518324 (01)00630414596143(10)N1518324(17)20290710; N1516224 (01)00630414596143(10)N1516224(17)20290621; N1515624 (01)00630414596143(10)N1515624(17)20290612; N1514924 (01)00630414596143(10)N1514924(17)20290605; N1514124 (01)00630414596143(10)N1514124(17)20290530; N1513324 (01)00630414596143(10)N1513324(17)20290522; N1512824 (01)00630414596143(10)N1512824(17)20290515; N1512424 (01)00630414596143(10)N1512424(17)20290510; N1511224 (01)00630414596143(10)N1511224(17)20290506; N1510724 (01)00630414596143(10)N1510724(17)20290423; N1510124 (01)00630414596143(10)N1510124(17)20290419.
Distribution pattern: US Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Congo, Republic C¿te dIvoire Croatia Cyprus Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kenya Kuwait Latvia Libya Malaysia Mexico Moldova, Rep. of Morocco Nepal Netherlands Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam
Derived failure modes
-
Manufacturing or process control
manufacturing defect