Device product
medex TranStar KIDS KIT 10IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9548
Z-0201-2025
Product summary
- Event
- Event 95438
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 3490 units
- Official record key
device-enforcement:Z-0201-2025
Official wording
Reason: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Code information: UDI/DI 50351688503006 (case), 10351688503008 (pouch), Lot Numbers: 4221225, 4250996, 4297929, 4304897,4316671, 4323138, 4328494, 4387189, 4404713
Distribution pattern: Worldwide distribution.
Derived failure modes
-
Manufacturing or process control
manufacturing defect