Skip to content
Recall Observatory FDA recall evidence

Device product

medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA, Product Code REF MXG300243R1

Z-0203-2025

September 24, 2024

Class II

Product summary

Firm
Smiths Medical ASD, Inc.
Event
Event 95438
Status
Ongoing
Classification
Class II
Quantity
3080 units
Official record key
device-enforcement:Z-0203-2025

Official wording

Reason: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Code information: UDI/DI 50351688518123 (case), 10351688518125 (pouch), Lot Numbers: 4282765, 4307302, 4309198, 4311461, 4316668, 4328487, 4350011, 4366881

Distribution pattern: Worldwide distribution.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing defect