Device product
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Z-0329-2025
Product summary
- Event
- Event 95490
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 77 units
- Official record key
device-enforcement:Z-0329-2025
Official wording
Reason: When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Code information: UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156
Distribution pattern: US Nationwide distribution in the states of NC.
Derived failure modes
-
Unknown
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.