Skip to content
Recall Observatory FDA recall evidence

Device product

medex TranStar MONITORING SET W/SAMPLING PORT 10/EA, Product Code REF MXG300260

Z-0204-2025

September 24, 2024

Class II

Product summary

Firm
Smiths Medical ASD, Inc.
Event
Event 95438
Status
Ongoing
Classification
Class II
Quantity
1640 units
Official record key
device-enforcement:Z-0204-2025

Official wording

Reason: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Code information: UDI/DI 50351688518833 (case), 10351688518835 (pouch), Lot Numbers: 4270120, 4297963, 4316640, 4319604, 4323118, 4287727, 4359814, 4387976, 4340473

Distribution pattern: Worldwide distribution.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing defect