Device product

NxTAG Respiratory Pathogen Panel, REF: I051C0447

Z-0042-2025

September 05, 2024

Class II

Product summary

Firm
Luminex Molecular Diagnostics Inc
Event
Event 95377
Status
Ongoing
Classification
Class II
Quantity
3375
Official record key
device-enforcement:Z-0042-2025

Official wording

Reason: Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.

Code information: UDI-DI: 00840487100417. Lot/Expiration: IK051C-0181/ 9/30/2024, IK051C-0182/ 9/30/2024, IK051C-0185/ 9/30/2024, IK051C-0186/ 9/30/2024, IK051C-0187/ 10/31/2024, IK051C-0188/ 10/31/2024, IK051C-0189/ 10/31/2024, IK051C-0190/ 10/31/2024, IK051C-0191/ 10/31/2024, IK051C-0192/ 10/31/2024, IK051C-0193/ 10/31/2024, IK051C-0194/ 10/31/2024, IK051C-0195/ 10/31/2024, IK051C-0196/ 11/30/2024, IK051C-0197/ 12/31/2024

Distribution pattern: U Nationwide distribution in the states of OH, CA, NC, WI, TN, TX, MI, NY, MO, FL, MA, NJ, PA, VA, AZ, IL, OK, CO, MN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.