Skip to content
Recall Observatory FDA recall evidence

Device product

CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24

Z-0014-2025

June 26, 2024

Class II

Product summary

Firm
Smiths Medical ASD Inc.
Event
Event 95305
Status
Ongoing
Classification
Class II
Quantity
41964 units
Official record key
device-enforcement:Z-0014-2025

Official wording

Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Code information: UDI/DI 10610586022111, Lot Numbers: 4040237, 4013368, 3983342, 4046817, 405240, 4060903

Distribution pattern: Worldwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.