Device product
Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782150 2) 782106
Z-3199-2024
Product summary
- Event
- Event 95289
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 10 units
- Official record key
device-enforcement:Z-3199-2024
Official wording
Reason: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Code information: Model UDI-DI: 782150 (01)00884838098329(21); 782106 (01)00884838098329(21) Serial Numbers: 26020 26023 26026 26034 46240 46270 46310 46315 46324 46352
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Derived failure modes
-
Unknown
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.