Device product
IntelliVue G7m Anesthesia Gas Module, Product Number 866173
Z-0007-2025
Product summary
- Event
- Event 95301
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 8,449 devices
- Official record key
device-enforcement:Z-0007-2025
Official wording
Reason: The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.
Code information: UDI-DI: 00884838051492; Serial Numbers: DE43800129 through DE43812627
Distribution pattern: US Nationwide. Global Distribution.
Derived failure modes
-
Unknown
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.