Skip to content
Recall Observatory FDA recall evidence

Device product

a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.

Z-0228-2025

September 18, 2024

Class II

Product summary

Firm
Baxter Healthcare Corporation
Event
Event 95402
Status
Ongoing
Classification
Class II
Quantity
2,967 beds
Official record key
device-enforcement:Z-0228-2025

Official wording

Reason: The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

Code information: All serial numbers with software version 1.27.000. UDI-DI 00887761985162: Product codes P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. UDI-DI 00887761999961: Product codes P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.

Distribution pattern: Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.