Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000, 8.3.1.000, 8.3.2.000, 8.4.0.000, 8.4.1.000 (including all SU s and HF s)

Z-0055-2025

September 06, 2024

Class II

Product summary

Firm
Agfa Healthcare NV
Event
Event 95411
Status
Ongoing
Classification
Class II
Quantity
801 units
Official record key
device-enforcement:Z-0055-2025

Official wording

Reason: Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.

Code information: UDI-DI: 05400874000710 Software Versions affected: All versions of Enterprise Imaging XERO Viewer are affected 8.1.2 or higher 8.1.4 or higher 8.2.0.100 or higher 8.2.2.000 or higher 8.3.2.000 or higher 8.4.0.000 or higher Enterprise Imaging XERO Viewer main version streams 8.4.0, 8.4.1, 8.3.2, 8.3.1, 8.3.0, 8.2.2, 8.2.1, 8.2.0, 8.1.5, 8.1.4, 8.1.3, 8.1.2, 8.1.1, 8.1, 8.0.1, 8.0.0, ICIS View 2014.1, ICIS View 2014.2

Distribution pattern: US Nationwide distribution. Foreign: Argentina, Australia, Belgium, Chile, Canada, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Georgia, Greece, Hong Kong, Iceland, Ireland, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Nicaragua, Norway, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Tanzania, Tunisia, United Arab Emirates, United Kingdom

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software defect