Device product
Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782151; 2) 782119; 3) 781358; 4) 782107; 5) 782136
Z-3200-2024
Product summary
- Event
- Event 95289
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 33 units
- Official record key
device-enforcement:Z-3200-2024
Official wording
Reason: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Code information: Model UDI-DI: 782151 N/A; 782119 (01)00884838104129(21); 781358 (01)00884838088115(21); 782107 (01)00884838098336(21); 782136 (01)00884838108608(21) Serial Numbers: 26500 26505 28502 28509 45393 45467 45599 62029 62041 62051 62052 62061 62066 62070 62071 62081 62082 62083 62091 62103 62111 62113 62114 62120 62126 62129 62147 62148 62186 62187 62188 66001 66002
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Derived failure modes
-
Unknown
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.